Split Double-sided Controlled 4 Glove Glove Box


This Split Double-sided controlled 4 glove Glove Box  are complete standalone systems integrated with entire functional components. They are able to create an inert environment with less than 1 ppm of H2O and O2. The systems are modulated with antechambers, removable windows, adjustable trays, lighting units, adjustable shelves, and gloves, which meet most of the operational needs in the glovebox. The systems are made of welded stainless steel and are equipped with highest-quality components. We also provide optional components to meet your special requirements.

Compare

Include:

1、Glovebox Chamber :1200mm(L) x 1000mm(W) x 900mm (H)
2、Single column Gas Purification System
3、Vacuum Pump
4、Gloves
5、Main Antechamber
6、Mini Antechamber
7、Oxygen Analyzer
8、Dew-point Analyzer
9、Solvent Purification System
10、Custom Logo

Category: Tag:
Brand Etelux
Product certification ISO9001 certification, CE certification, UL certification
Structure integrated structure
Power supply 110-220V AC,50/60Hz, single-phase
Box Box Material 304 Stainless Steel (Type 304),

Thickness: 3mm

Box Size: 1200mm(L) x 1000mm(W) x 900mm(H)

47.2″ x 39.3″ x 35.4″

Box Support height 900mm(35.43″)
Front window 8mm thickness, sealing ring using 3/8 inch thick OMEGA sealing ring
Environment of Box inside H2O <1 ppm (20°C/68 ℉, 1 atm)

O2 <1 ppm (20°C/68 ℉, 1 atm)

Antechamber Control way manual
Main chamber size Φ360mm (Dia.)x 600mm(length)

14.17 “(Dia.) x 23.62” (L)

Mini chamber size Φ150 mm(Dia.) x 300mm(L)

5.9 “(Dia.) x11.8” (L)

Vacuum degree <=-0.1MPa
Working gas 99.999% high pure N2, Ar, He and other inert gas
Control gas:99.999% high pure inert gas
Regeneration gas: mixing gas of working gas and H2,hydrogen(H2)Volume accounted for 5-10%
Gas purification system H2O<1ppm, O2<1ppm
Pressure control system be auto controlled by PLC touch screen/ foot switch
Control system SIEMENS colors touch screen (6 inches), PLC control system, Bilingual in Chinese and English can be switched
Pressure sensor -2500~2500Pa touch screen, accuracy±1Pa
Pressure control control the box pressure to one setting point (scope can be set)
measurements Moisture analyzer US GE brand, 0~1000ppm touch screen display, precision 0.1ppm
Oxygen analyzer US GE brand, 0~1000ppm touch screen display, precision 0.1ppm
Vacuum Pump EDWARDS, pumping speed: 8.4-12m3/h, Model: RV12
Glove Material Butyl
Thickness 0.4mm or 0.8mm (available for selection)
Optional equipment Refrigerator temperature -35 ℃, volume 18L or 32L optional;
Heating chamber The main antechamber is equipped with a heating system with a temperature of 200 ℃ and a temperature control of ± 1 ℃