- The Michell Instruments XTP601 Oxygen Analyzer is a thermo-paramagnetic sensor that is a robust, linear and stable oxygen analyzer for measurements in gases such as hydrogen, nitrogen or carbon dioxide. The sensor is housed in an IP66 rugged casing, making it suitable for a wide range of applications. When supplied with flame arrestors, the analyzer becomes explosion-proof and suitable for use with flammable gases and/or installation in hazardous areas
- The XTP601 has barometric pressure compensation fitted as standard and is ranged specifically to customer’s needs to offer the best accuracy where it matters.
- The sensor in Michell’s XTP601 analyzer largely eliminates thermistor drift associated with other comparable devices, improving the stability of the measurement. This allows longer calibration intervals and reduces both labor and consumable costs.
Michell Instruments XTP601 Oxygen Analyzer
Features
- Available in 3 models:
- General purpose analyzer
- General purpose analyzer with flame arrestors
- Hazardous area analyzer
- ATEX, IECEx, CCSAUS, TC, TR & Ex hazardous area certified
- Touch-screen display
- Thermo-paramagnetic sensor
- RS485 and two 4 to 20mA outputs
- Accuracy 1% full scale
- Zero range stability better than 0.25% of range per month
- Ranges available: 0 to 0.5% up to 0 to 50% & 20, 80 or 90–100%
- Backlit LCD display
- Accuracy less than ±1% of span or ±0.02% O2
- Meets the requirements of IEC61508 SIL 2
$5,750.00 $6,000.00
Measurement technology | Thermo-paramagnetic oxygen sensor |
---|---|
Gas | Process and non-condensing sample with particles <3μm |
Measurement range | Selectable from 0–0.5% up to 0-50% and 20, 80 or 90–100% |
Display resolution | 0.01% for suppressed zero ranges |
Display type | Backlit LCD |
Accuracy (excluding suppressed zero ranges) | <±1% of span or ±0.02% O2 whichever is greater |
Accuracy for suppressed zero ranges (80/90–100%) | <0.2% O2 |
Response time (T90) with high speed response (HSR) enabled | <15 seconds |
Repeatability | ±0.2% of span or 0.01% O2 whichever is greater |
Linearity | ±0.5% of span or 0.05% O2 whichever is greater |
Zero stability | ±0.25% of span per month |
Span stability | ±0.25% of span per month |
Sample flow rate | GP1 and GP3: 100 to 500 ml/min (0.2 to 1.06 scfh) |
Sample pressure | 0 to 3 barg inlet with atmospheric vent |
Sample temperature | 5 to 55°C (+41 to +131°F) max (+45°C (+113°F) for standard cell) |
Sample cell temperature | Standard +50°C (+122°F)(Optional +55°C and +60°C (+131°F and +140°F) for ranges 0-0.5% up to 0-50% only.) |
Background gas | Analyzer is calibrated in the background gas of the process. Standard gases are: N2, CO2, CH4, H2 & Biogas. Others available on request. |
Analog inputs | 2 off 4–20 mA inputs |
Analog outputs | 2 off 4–20 mA outputs (powered with 24V excitation voltage) |
Output ranges | Primary range is set to the calibrated range of the instrument |
Alarms | 2 off single pole changeover (SPCO) relays for O2 concentration (250 V, 5 A max) |
Datalogging | The user can use the digital communications to log the output from the analyzer The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp |
Digital communications | Modbus RTU over RS485 Protocol |
Power supply | 24 V DC; 1.5 A max |
Cable entries | Analyzer is supplied with 3 x M20 cable entries. Cable glands, conduit entries and blanking plugs are available as accessories. |
Ambient temperature | 5 to +55°C (+41 to +131°F) |
Atmopheric pressure | 750mbar to 1250mbar |
Warm up time | <25 minutes |
Stabilization time | 5 minutes |
Dimensions | 234 x 234 x 172mm (9.2 x 9.2 x 6.7”) (w x d x h) |
Weight | 9.7kg (21.4lbs) |
Wetted materials | 316 & 430F stainless steel, boroscillicate glass, platinum, 3M 2216 (plus o-ring) |
O-Ring materials | Viton, Silicone or Ekraz |
Gas connection | EX1, EX3 & GP2: 1/8” NPT (female) |
Ingress protection | IP66, NEMA 4X |
ATEX | II 2GD Ex d IIB+H2 T6 Gb |
IECEx | Ex d IIB+H2 T6 Gb |
CCSAus | Class I, Division 1, Groups B,C,D, T6 |
INMETRO | Ex d IIB+H2 T6 Gb |
TC TR Ex | 1 Ex d IIB+H2 T6 Gb |
Temperature ranges for ATEX and IECEx as per o-ring type | Silicon: Ta = -40°C to +55°C |