Michell Instruments XTP601 Oxygen Analyzer


Features

  • Available in 3 models:
    • General purpose analyzer
    • General purpose analyzer with flame arrestors
    • Hazardous area analyzer
  • ATEX, IECEx, CCSAUS, TC, TR & Ex hazardous area certified
  • Touch-screen display
  • Thermo-paramagnetic sensor
  • RS485 and two 4 to 20mA outputs
  • Accuracy 1% full scale
  • Zero range stability better than 0.25% of range per month
  • Ranges available: 0 to 0.5% up to 0 to 50% & 20, 80 or 90–100%
  • Backlit LCD display
  • Accuracy less than ±1% of span or ±0.02% O2
  • Meets the requirements of IEC61508 SIL 2

$5,750.00 $6,000.00

Compare
  • The Michell Instruments XTP601 Oxygen Analyzer is a thermo-paramagnetic sensor that is a robust, linear and stable oxygen analyzer for measurements in gases such as hydrogen, nitrogen or carbon dioxide. The sensor is housed in an IP66 rugged casing, making it suitable for a wide range of applications. When supplied with flame arrestors, the analyzer becomes explosion-proof and suitable for use with flammable gases and/or installation in hazardous areas
  • The XTP601 has barometric pressure compensation fitted as standard and is ranged specifically to customer’s needs to offer the best accuracy where it matters.
  • The sensor in Michell’s XTP601 analyzer largely eliminates thermistor drift associated with other comparable devices, improving the stability of the measurement. This allows longer calibration intervals and reduces both labor and consumable costs.
Category:
Measurement technology

Thermo-paramagnetic oxygen sensor

Gas

Process and non-condensing sample with particles <3μm

Measurement range

Selectable from 0–0.5% up to 0-50% and 20, 80 or 90–100%

Display resolution

0.01% for suppressed zero ranges

Display type

Backlit LCD

Accuracy (excluding suppressed zero ranges)

<±1% of span or ±0.02% O2 whichever is greater

Accuracy for suppressed zero ranges (80/90–100%)

<0.2% O2

Response time (T90) with high speed response (HSR) enabled

<15 seconds

Repeatability

±0.2% of span or 0.01% O2 whichever is greater

Linearity

±0.5% of span or 0.05% O2 whichever is greater

Zero stability

±0.25% of span per month

Span stability

±0.25% of span per month

Sample flow rate

GP1 and GP3: 100 to 500 ml/min (0.2 to 1.06 scfh)
EX1, EX3 and GP2: 270 to 330 ml/min (0.6 to 0.7 scfh)

Sample pressure

0 to 3 barg inlet with atmospheric vent

Sample temperature

5 to 55°C (+41 to +131°F) max (+45°C (+113°F) for standard cell)

Sample cell temperature

Standard +50°C (+122°F)(Optional +55°C and +60°C (+131°F and +140°F) for ranges 0-0.5% up to 0-50% only.)

Background gas

Analyzer is calibrated in the background gas of the process. Standard gases are: N2, CO2, CH4, H2 & Biogas. Others available on request.

Analog inputs

2 off 4–20 mA inputs
One for an external sensor that can be displayed on the screen
One to act as an active compensation for the process conditions

Analog outputs

2 off 4–20 mA outputs (powered with 24V excitation voltage)

Output ranges

Primary range is set to the calibrated range of the instrument
The second is user selectable 0-100%

Alarms

2 off single pole changeover (SPCO) relays for O2 concentration (250 V, 5 A max)

Datalogging

The user can use the digital communications to log the output from the analyzer The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp

Digital communications

Modbus RTU over RS485 Protocol

Power supply

24 V DC; 1.5 A max

Cable entries

Analyzer is supplied with 3 x M20 cable entries. Cable glands, conduit entries and blanking plugs are available as accessories.

Ambient temperature

5 to +55°C (+41 to +131°F)

Atmopheric pressure

750mbar to 1250mbar

Warm up time

<25 minutes

Stabilization time

5 minutes

Dimensions

234 x 234 x 172mm (9.2 x 9.2 x 6.7”) (w x d x h)

Weight

9.7kg (21.4lbs)

Wetted materials

316 & 430F stainless steel, boroscillicate glass, platinum, 3M 2216 (plus o-ring)

O-Ring materials

Viton, Silicone or Ekraz

Gas connection

EX1, EX3 & GP2: 1/8” NPT (female)
GP1 and GP3: 1/4” NPT (female)

Ingress protection

IP66, NEMA 4X

ATEX

II 2GD Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66

IECEx

Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66

CCSAus

Class I, Division 1, Groups B,C,D, T6

INMETRO

Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66

TC TR Ex

1 Ex d IIB+H2 T6 Gb

Temperature ranges for ATEX and IECEx as per o-ring type

Silicon: Ta = -40°C to +55°C
Viton: Ta = -15°C to +55°C
Ekraz: Ta = -10°C to +55°C