Process
Gas circulation: Closed loop gas recirculation
Gas Purification System
Removal of H2O and O2
Working Gas
Inert gas: Nitrogen and Argon
Attainable Purity
H2O < 1 ppm, O2 < 1 ppm
Purifier
Amount/type:Double H2O/O2 purifier column
Capacity: Oxygen removal: 30 l (standard conditions), Moisture removal: 1600 g
Lindy Molecular Sieve,and BASF R3-11 Copper Catalyst.
Purification Columns Include Fill & Empty Ports for Easy Catalyst Replacement
Material: Stainless Steel type 1.4301 (US type 304)
Heater: Integrated
Circulation Unit
Type: Integrated blower vacuum-tight, oil-free
Flow rate: 100 m³/h (120 cfm) at ΔP = 60 mbar (60 Hz)
Features: Operated with all standard power supplies; optimum system tuning by frequency preselection for the blower
Valves
Main valves: Stainless DN 40 Valves
Control valves: magnetic valve system, DN 6/10 (Brass)
Piping
Main piping: Stainless pipe DN 40 KF-flange system*
Control pipework: Copper pipe DN 6/10*
*Stainless Steel pipe 1.4301 (US type 304) on request
Integral Leak Rate
Less than 10-5 mbar l/s