AII GPR-1500 GB / GPR-2500 GB Oxygen Monitors


Features

  • Measurement ranges
    • GPR-1500 GB: 0-10, 0-100, 0-1000 ppm; 0-1%, 0-25%
    • GPR-2500 GB: 0-1, 0-5, 0-10, 0-25%
  • Sensitivity
    • GPR-1500 GB: 0.05 ppm
    • GPR-2500 GB: 0.005%
  • <2% of selected range accuracy
  • <1% of scale linearity
  • 0.34-2 barg (5-30 psig) with atmospheric vent inlet pressure
  • 0.5-1 Nl/min (1-2 SCFH) flow rate

$1,525.00$2,698.00

Compare
  • The AII GPR-1500 GB and GPR-2500 GB oxygen monitors measure oxygen in glove boxes ranging from air (21% O2) down to low levels (0-10 ppm O2).
  • The GPR-1500 GB is used for trace oxygen measurements as low as 0.1 ppm oxygen in various background gases. The sensor is housed in a stainless steel case that can be screwed into place.
  • The GPR-2500 GB is used for oxygen measurements from 21% down to 100 ppm oxygen in a range of gas backgrounds. The analyzer or remote sensors can be connected directly on the glove box using a KF-40 flange. In the remote version, the oxygen sensor can be placed in the glove box using the sample/calibration module.
  • The sensors are designed to avoid potential weaknesses common in typical galvanic cell design. Each sensor type has been specifically engineered to provide the optimum balance between performance and longevity for individual applications. In the absence of oxygen, the sensor will produce zero output. The sensor is linear up to 100%, and in most cases only a span calibration is required.
  • For applications with a background gas containing more than 0.5% CO2, the specially designed XLT sensor should be selected. The sensor has a special electrolyte formula that can operate in temperatures as low as -10°C.
SKU: N/A Category:
GPR-1500 GB GPR-2500 GB
Measurement range 0-10, 0-100, 0-1000 ppm, 0-1%, 0-25% 0-1%, 0-5%, 0-10%, 0-25%
Accuracy < 2% of selected range at constant conditions < 2% of selected range at constant conditions
Response time T90 < 10 seconds T90 < 10 seconds
Recovery time 60 seconds in air to < 10 ppm in < 1 hour on N2  purge Not applicable
Sensitivity (LDL) 0.05 ppm 0.005 %
Linearity < 1% of scale < 1% of scale
Sensor model GPR-12-333-M GPR-11-32-4
XLT-12-333-M for gases containing > 0.5% CO2 XLT-11-24-4 for gases containing > 0.5% CO2
Sensor life at 25ºC (77°F) and 1 atm 24 months in < 1000 ppm O2
6 months in air
GPR-11-32-4 32 months in air
XLT-11-24-4 24 months in air
Calibration interval 30 days
Inlet pressure Nominally atmospheric
Flow through system: 0.34-2 barg (5-30 psig) with atmospheric vent
Flow rate Ambient monitoring or
Flow through system 0.5-1 Nl/min (1-2 SCFH)
Gas connections KF 40 flange or
1/8” compression tube fittings (with flow through housing)
Wetted parts GPR-1500 GB Stainless steel GPR-2500 GB Delrin
Display Graphical LCD 7 x 3.5cm (2.75 x 1.375”); resolution 0.001
Enclosure Painted aluminum, 7.6 x 10.1 x 5.1cm (3 x4 x2”)
Weight 900g (2lbs)
Compensation Temperature
Signal output 4-20mA (loop current)
Alarms None
Operating temperature GPR sensor: 5ºC to 45ºC (41°F to 113°F) XLT sensor: -10ºC to 45ºC (14°F to 113°F)
Power 18-24 V DC two wire loop
Area classification General purpose