-13%

NEW Agilent Varian IDP-07 IDP07 IDP7 Dry Scroll Vacuum Pump, 1-Phase, PN X3807-64000


$4,747.50 $5,486.00

Compare

The new Agilent Varian IDP-07 dry scroll vacuum pump is designed for extremely quiet < 55 dBA and low vibration operation, delivering a peak pumping speed of 5.4 cfm (9.1 m3/h) at 60 Hz, and produces an ultimate pressure of 20 mTorr. These new Agilent IDP-07 dry scroll pumps are hermetically-sealed and are appropriate for helium recirculation as well as vacuum oven, and backing pump applications. The hermetic design provides a clean gas path through the pump, which eliminates any risk of oil or grease contamination allowing the motor and bearings to run at atmospheric pressure, isolating them from exposure to vacuum process gases or water vapor. The Agilent IDP-07 supports a global single-phase motor for operating voltages ranges easily switched between 100-120V or 200-230 VAC at 50/60Hz.

This IDP-07 model, Agilent technologies part number X3807-64000, DOES NOT include the optional Integrated Inlet Isolation Valve which protects against backward migration of air through the scroll pump inlet if the IDP-07 loses power. This isolation valve prevents sudden venting of customer’s chamber or instrument on system power loss.

High vacuum pumps, such as turbo molecular pumps, are increasingly used in conjunction with dry backing pump technology. Pumps based on scroll technology have become the dry pump technology of choice for backing turbo molecular high vacuum pumps. There are several factors that make these Agilent Varian IDP-07 scroll technology the superior dry pump choice, including clean, predictable performance, lower vibration and noise, and long service life.

Modern Agilent IDP-07 scroll technology is based on smooth orbital operation; the vacuum pump cycle of gas compression and discharge occurs with minimal pulsation and very low vibration. Depending on the application, vibration can have adverse effects, and may necessitate the use of additional components such as vibration isolators. The new Agilent IDP-07 is the dry pump of choice in a wide variety of applications where excessive vibration is unacceptable.

Current scroll pump technology and tip seal design are factors that work together to provide low base pressure, and a gradual increase in base pressure over the service life. This gradual increase in base pressure allows maintenance to be done before the vacuum levels become unacceptable and cuts unscheduled downtime. A typical scroll pump like the Agilent IDP-07 will achieve a base pressure of 20 mTorr.

These Agilent IDP-07 dry scroll pumps create vacuum using a simple dual scroll mechanism in which one of the nested scrolls orbits about the other, creating moving zones of captured gas. Gas enters the scroll set at the outside perimeter and is displaced and compressed toward the center of the hub where it is exhausted. The uncomplicated dual scroll design offers many benefits including lower noise and vibration levels, simple and infrequent maintenance requirements, plus the elimination of catastrophic failure modes.

These new Agilent Technologies IDP-07 includes improved solid tip seals for longer tip seal life and longer service interval. The solid tip seal material performs well for condensable vapor applications. The tip seal replacement for the IDP-07 is easier than ever before. In addition to a tip seal kit (X3807-67000), a few simple tools are all that is required to replace a worn tip seal: a flat head screw driver, 4 and 5 mm Allen wrenches, isopropyl alcohol, a few lint-free wipes. Please see pages 18 – 20 of the instruction manual for tip seal replacement procedure. An instruction manual and brochures in PDF format for the new Agilent IDP-07 hermetically-sealed dry scroll vacuum pumps can be downloaded in our available download section below.

Category:
IDP-7 dry scroll single sided hermetic vacuum pump
Peak pumping speed 60 Hz: 152 L/min (9.1 m3/h); 50 Hz: 120 L/min (7.2 m3/h)
Media No corrosive, explosive or particulate forming gases
Ultimate pressure 60 Hz: 2.6 x 10-2 mbar (2.0 x 10-2 Torr)
50 Hz: 4.0 x 10-2 mbar (3.0 x 10-2 Torr)
Max. inlet pressure 1.0 atmosphere (0 psig)
Max. recommended continuous inlet pressure 200 mbar (150 Torr)
Max. outlet pressure at nom voltage 1.34 atmosphere (5 psig)
Inlet connection NW25
Exhaust connection NW16
Gas ballast Female, 0.25 inch, national pipe thread, (20 µm sintered plug provided)
SAE-2 5/16-24
Water vapor handling per Pneurop 6602 120 grams/hr maximum
Ambient operating temperature 5 to 40 °C (41 to 104 °F)
Storage temperature –20 to 60 °C (–4 to 140 °F)
Motor rating 0.38 HP (300 W)
Operating voltages 100-120; 200-240 VAC, 50/60 Hz, single phase, 3 to 6A typical
Motor thermal protection Automatic thermal protection
Operating speed 60 Hz: 1750 rpm; 50 Hz: 1450 rpm
Cooling system Air-cooled
Weight Pump only: 24.5 kg (54 lbs), shipping weight: 28.1 kg (62 lbs)
Dimensions (H x W x L mm) 254 x 260 x 420
Leak rate (with exhaust sealed) <1 x 10-6 sccs (1 x 10-6 mbar.l/s)
Noise level 52 +/- 2 dB (A)
Vibration level at inlet 7 mms-1 rms
Hour meter Integral hour meter provided
Emissions and immunity EN 61236-1:2013 as a Class A device
Safety EN/IEC 61010 3rd Edition certified