AII GPR-2900 Oxygen Analyzer


Features

  • Measuring range:
    • 0-1%, 0-5%, 0-10%, 0-25%;
    • Optional 0-1%, 0-5%, 0-10%, 0-100%
  • Accuracy: <2%
  • Barometric pressure and temperature compensation
  • Two user configurable alarms
  • Range ID output
  • 12 to 28 V DC or mains powered options
  • XLT sensor options for CO2 backgrounds

$1,658.00$3,647.00

Compare
  • The AII GPR-2900 Oxygen Analyzer is a general purpose oxygen analyzer with barometric pressure and temperature compensation, four measurement ranges, and two user configurable alarm contacts. All versions can be locked manually to a single range or auto-range. The analyzer also features a 4-20mA isolated signal output, 4-20mA or 1-5V DC range ID outputs, Delrin remote sensor housing with 1/8-inch tube fittings, AC or DC power, and a panel mount aluminum enclosure.
  • The GPR-2900 operates with a remote sensor to allow flexibility of installation, and measures O2 up to 100%. The remote sensor block allows an easy oxygen sensor replacement for simple and fast maintenance and low cost of ownership.
SKU: N/A Category:
Measurement range 0-1%, 0-5%, 0-10%, 0-25%;
optional 0-1%, 0-5%, 0-10%, 0-100%
Accuracy < 2% of selected range at constant conditions
Response time T90 < 10 seconds
Recovery time NA
Sensitivity (LDL) 0.005%
Linearity < 0.5% of scale
Sensor model GPR-11-60-4
XLT-11-24-4 for gases containing > 0.5% CO2
Sensor life at 25ºC (77°F) and 1 atm GPR-11-60-4  60 months in air XLT-11-24-4  24 months in air
Calibration interval 30 days
Inlet pressure 0.34 2 barg (5-30 psig) with atmospheric vent
Flow rate 0.5 1.0 l/m (1-2 SCFH)
Gas connections 1/8” compression tube fittings
Wetted parts Delrin; stainless steel optional
Display Graphical LCD 12.7 x 7cm (5 x 2.75”)
Enclosure Painted sheet metal 6” x 4” x 4” panel mount
Compensation Barometric pressure and temperature
Signal output 4-20 mA or 1-5V
Range ID 4-20 mA non-isolated or 1-5V
Option: reassign relay contacts as range ID (eliminates standard alarms).
Alarms Two user configurable alarms: magnetic coil relays rated 3A at 100 V AC
Operating temperature GPR sensors: 5ºC to 45ºC (41°F to 113°F)
XLT sensors: -10ºC to 45ºC (14°F to 113°F)
Power 100-240 V AC or 12-28 V DC