AII GPR-1000 Oxygen Analyzer


Features

  • Measuring range: 0-1000 ppm, 0-1%, 0-25%
  • Accuracy <2%
  • Sensor models:
    • GPR-12-100-M
    • XLT-12-100-M
  • 24 to 32 months sensor life span (in normal use)
  • Internal pump option
  • General purpose or intrinsically safe (ATEX) models
  • 0 to 1V output
  • 0.5-1.0 Nl/m (1-2 SCFH) flow rate
  • XLT sensor options for CO2 backgrounds
  • Lightweight: 2.5Kgs

$3,009.00$3,825.00

Compare
  • The AII GPR-1000 Oxygen Analyzers is designed for measuring oxygen below 1,000ppm with an LDL of 5ppm and is temperature compensated.
  • The GPR-1000 oxygen meter can also be operated up to 1% range.
  • The GPR-1000 is supplied with brass fittings. Available options for the GPR-1000 include XLT sensor backgrounds, stainless steel fittings, weld purge kit.
  • Ranges for for the GPR-1000 include 0 to 1,000ppm & 0 to 1% (0 to 25% for calibration only).
SKU: N/A Category:
Measurement range 0-1000 ppm, 0-1%, 0-25%
Accuracy < 2% of selected range at constant conditions
Response time T90 < 10 seconds
Recovery time 60 sec in air to < 100 ppm in < 15 min on N2 purge
Sensitivity (LDL) 5 ppm
Linearity < 1% of scale
Sensor model GPR-12-100-M
XLT-12-100-M for gas mixture with > 0.5% CO2
Sensor life at 25ºC (77°F) and 1 atm GPR-11-32-RTS 32 months in air
XLT-11-24-RTS 24 months in air
Calibration interval 30 days
Inlet pressure 0.34–2 barg (5-30 psig) with atmospheric vent
Flow rate 0.5-1.0 Nl/m (1-2 SCFH)
Gas connections 1/8″ quick connect WP 1/8” compression fittings
Wetted parts Brass connections; stainless steel optional
Display Graphical LCD 7 x 3.5cm (2.75 x 1.375”)
Resolution 1 ppm
Enclosure Painted aluminum NEMA 4X, 10.1 x 22.9 x 7.6cm (4 x 9 x 3”)
3.64Kg (8lbs)
Compensation Barometric pressure and temperature
Signal output 0-1V
LED indicators LOW BATT (72 hr. warning); CHARGE mode
Operating temperature 5°C to 45°C (41°F to 113°F) (GPR sensor),
-10° to 45°C (14°F to 113°F) (XLT sensor)
Power Rechargeable battery (lead acid)
Battery life Up to 30 days on a single charge (40 hours with pump running)
Area classification ATEX:  Ex ib IIB T4 Tamb    +5°C to +45°C