- The AII Palm O2 oxygen analyzer is a simple and cost-effective choice to check for oxygen or carbon dioxide in compressed air breathing tanks intended for scuba diving. With only two buttons and one-touch calibration, the Palm O2 is easy to use. The analyzer offers a battery life of up to 18 months of continuous use, and is available with internal or external sensor.
- The Palm O2 uses an electrochemical galvanic fuel cell-type oxygen sensor of the type that is extensively used to measure oxygen concentrations from 0% to 100% in gas streams. A battery-powered microprocessor converts the sensor’s signal output representing the partial pressure of oxygen in the gas stream being analyzed. The resulting oxygen reading is displayed by a large easy-to-read backlit liquid crystal display (LCD) that has a resolution of 0.1% oxygen. The sensor also has an absolute zero reading that displays 0.00 oxygen when no oxygen is present. The microprocessor is controlled from a keypad and provides features like system diagnostics, warning indicators, controls and an alarm capability for continuous monitoring that enhance both safety and effectiveness.
- The analyzer is designed to be handheld, but can be temporarily placed in a fixed position on a pole or shelf with optional bracket attachments. A remote sensor option is available, which makes the Palm O2 more flexible and easier to use.
Oxygen Analyzers
AII Palm O2 Oxygen Analyzer
Features
- 0-100% O2 measurement range
- ±2% accuracy
- 0.50% sensitivity (LDL)
- ±1% linearity under constant conditions
- Ambient or 0.34-2 barg (5-30 psig) inlet pressure
- 0.2-10 Nl/min (0.4-20 SCFH) flow rate
- Internal or external sensor
$270.00 – $306.00
CompareMeasurement range | 0-100% oxygen |
Accuracy | < ±2% of FS range under constant conditions |
Response time | T90 < 10 seconds |
Sensitivity (LDL) | 0.50% |
Linearity | ±1% under constant conditions |
Sensor model | AII-11-75-PO2 AII-11-75-PO2-D (for diving) |
Sensor life at 25°C (77°F) and 1 atmosphere | 32 months in air |
Calibration interval | Prior to use and after 8 hours operation |
Inlet pressure | Ambient or 0.34-2 barg (5-30 psig) for flow through connector |
Flow rate (constant) | 0.2-10 Nl/min (0.4-20 SCFH) |
Gas connections | 1×16 mm thread 1/4” tube for remote sensors See accessories for additional connectors |
Display | 3 digit LCD 2.8 x 1.6cm (1.1 x 0.625”); resolution 0.1 |
Dimensions | 6.9 x 10.4 x 3.4cm (2.72” x 4.1” x 1.35”) |
Weight | 196g (7oz) |
Compensation | Temperature compensated |
Alarms | User adjustable HI and LO alarms (AII-3000 M and MHC only) |
Storage temperature | -20°C to +60°C (-4°F to 140°F) on intermittent basis |
Operating temperature | 5 to 45°C (41°F to 113°F) |
Power | 2 AA alkaline batteries |
Battery Life | Up to 13,000 hrs |
Humidity | Non-condensing 0-95% RH |
Warm up time | None |
Approvals | ISO 9001:2000, MDD 93/42/Annex II, ISO 13485:2003 |