-10%

AII Palm O2 Oxygen Analyzer


Features

  • 0-100% O2 measurement range
  • ±2% accuracy
  • 0.50% sensitivity (LDL)
  • ±1% linearity under constant conditions
  • Ambient or 0.34-2 barg (5-30 psig) inlet pressure
  • 0.2-10 Nl/min (0.4-20 SCFH) flow rate
  • Internal or external sensor

$270.00$306.00

Compare
  • The AII Palm O2 oxygen analyzer is a simple and cost-effective choice to check for oxygen or carbon dioxide in compressed air breathing tanks intended for scuba diving. With only two buttons and one-touch calibration, the Palm O2 is easy to use. The analyzer offers a battery life of up to 18 months of continuous use, and is available with internal or external sensor.
  • The Palm O2 uses an electrochemical galvanic fuel cell-type oxygen sensor of the type that is extensively used to measure oxygen concentrations from 0% to 100% in gas streams. A battery-powered microprocessor converts the sensor’s signal output representing the partial pressure of oxygen in the gas stream being analyzed. The resulting oxygen reading is displayed by a large easy-to-read backlit liquid crystal display (LCD) that has a resolution of 0.1% oxygen. The sensor also has an absolute zero reading that displays 0.00 oxygen when no oxygen is present. The microprocessor is controlled from a keypad and provides features like system diagnostics, warning indicators, controls and an alarm capability for continuous monitoring that enhance both safety and effectiveness.
  • The analyzer is designed to be handheld, but can be temporarily placed in a fixed position on a pole or shelf with optional bracket attachments. A remote sensor option is available, which makes the Palm O2 more flexible and easier to use.
SKU: N/A Category:
Measurement range 0-100% oxygen
Accuracy < ±2% of FS range under constant conditions
Response time T90 < 10 seconds
Sensitivity (LDL) 0.50%
Linearity ±1% under constant conditions
Sensor model AII-11-75-PO2
AII-11-75-PO2-D (for diving)
Sensor life at 25°C (77°F) and 1 atmosphere 32 months in air
Calibration interval Prior to use and after 8 hours operation
Inlet pressure Ambient or 0.34-2 barg (5-30 psig) for flow through connector
Flow rate (constant) 0.2-10 Nl/min (0.4-20 SCFH)
Gas connections 1×16 mm thread
1/4” tube for remote sensors
See accessories for additional connectors
Display 3 digit LCD 2.8 x 1.6cm (1.1 x 0.625”); resolution 0.1
Dimensions 6.9 x 10.4 x 3.4cm (2.72” x 4.1” x 1.35”)
Weight 196g (7oz)
Compensation Temperature compensated
Alarms User adjustable HI and LO alarms (AII-3000 M and MHC only)
Storage temperature -20°C to +60°C (-4°F to 140°F) on intermittent basis
Operating temperature 5 to 45°C (41°F to 113°F)
Power 2 AA alkaline batteries
Battery Life Up to 13,000 hrs
Humidity Non-condensing 0-95% RH
Warm up time None
Approvals ISO 9001:2000, MDD 93/42/Annex II, ISO 13485:2003