-6%

Panametrics O2.IQ Oxygen Analyzer Package


Features

  • ±1% accuracy of range at calibration point
  • oxy.IQ electrochemical analyzer
  • 32 to 113ºF (0 to 45ºC) operating & sample gas temperature
  • 0 to 400 psig (0 to27.5 Bar) sample gas pressure
  • 24 to 28 VDC at 50mA required power
  • 1/8 in NPT inlet and outlet process connection
  • 1.0 SCFH (500 cc/min) sample flow rate
  • O2 sensors ordered separately

$5,150.00$6,150.00

Compare
  • The Panametrics O2.IQ Oxygen Analyzer Package simplifies and standardizes your options for reliable oxygen measurement in a variety of applications. At the heart of this package is the oxy.IQ analyzer supported by a sample conditioning system purposefully designed to maintain ease of access and functionality in a space conscious housing. Simply mount the O2.IQ, connect to the two terminals and the analyzer is ready for its field calibration.
  • The standard O2.IQ package has been designed with oxygen measurement in natural gas pipelines in mind; however, the package is easily transferable to a wide range of applications, given its standard sampling conditioning system.The pressure regulation functionality allows for inlet pressures of up to 400 psig (27.5 bar) and reduces to ambient pressure for accurate measurement. The compact footprint of this stainless-steel package combined with the loop powered oxy.IQ analyzer make it perfect for safe and hazardous area locations.

 

SKU: N/A Category:
Accuracy

• ±1% of range at calibration point
• ±2% of range at the calibration point for the 0 to 10 ppmv range (OX-1 or OX-2 only)

Repeatability

• ±1% of range
• ±2% of range for the 0 to 10 ppmv range (OX-1, OX-2 only)

Resolution

• ±0.1% of range

Linearity

• ±2% of range (OX-1, 2, 3, 5)
• ±5% of range (OX-4)

O2 Sensor operating temperature

32 to 113ºF (0 to 45ºC)

Sample pressure

Vented to atmosphere during operation and calibration

Atmospheric pressure effect

±0.13% of reading per mmHg (directly proportional to absolute pressure). During calibration, pressure and flow must be kept constant.

Process connection

1/8 in. NPT inlet and outlet

Dimensions

4.10 x 2.75 x 2.05 in (104.1 x 69.9 x 52.1 mm)

Weight

1.35 lb (612 grams)

Sample flow rate

1.0 SCFH (500 cc/min) recommended for process units

Electrical classification certification

Intrinsically safe
USA/Canada
IS for Class 1, Groups ABCD, T4
AEx ia IIC T4
ATEX and IECEx
Ex ia IIC Ga T4 Tamb -20 to 60ºC
IS package, non-incendive without use of Zener barrier or galvanic isolator:
• USA/Canada: Class 1, Div 2, Groups A, B, C, D: T4

Approvals

GP – General Purpose, CL1-DIV2 – Class 1 Div 2 – Non incendive, CL1-DIV1 – Class 1 Div 1/ATEX_IECEX IS, IS – Intrinsically Safe

Barrier Type

None, -Z – Zener, -G – Galvanic

Units

-I – Imperial, -M – Metric, Imperial, Metric